Gamma spectrum and coincidence summation simulations with Geant4 in the analysis of radionuclide using BEGe detector

Simple item page

Simple item page

Full item details

dc.contributor.author
Liu, Chuanlei
Zhang, Weihua
Pierce, Dylan
Ungar, Kurt
Tulk, Christopher
Bean, Marc
dc.date.accessioned
2024-04-19T17:30:36Z
dc.date.available
2024-04-19T17:30:36Z
dc.date.issued
2018-04-05
dc.description.abstract - en
A Geant4 application has been developed to simulate the gamma spectrum of a BEGe detector. The summation effects were encompassed for both random and true coincidences through a time-based approach. The application well reproduces the typical spectral features such as full energy peaks, escape peaks, summation peaks and Compton/backscatter continuum. The simulated gamma response is consistent with empirical data within 3%, while simulations of the X-ray and baseline regions are valid within a 10% deviation.
dc.description.abstract-fosrctranslation - fr
Une application Geant4 a été développée pour simuler le spectre gamma d'un détecteur BEGe. Les effets de sommation ont été pris en compte à la fois pour les coïncidences aléatoires et réelles grâce à une approche basée sur le temps. L'application reproduit bien les caractéristiques spectrales typiques telles que les pics de pleine énergie, les pics d'échappement, les pics de sommation et le continuum Compton/rétrodiffusion. La réponse gamma simulée est cohérente avec les données empiriques à 3 % près, tandis que les simulations des régions de rayons X et de référence sont valides à un écart de 10 %.
dc.identifier.doi
https://doi.org/10.1016/j.apradiso.2018.04.003
dc.identifier.uri
https://open-science.canada.ca/handle/123456789/2355
dc.language.iso
en
dc.publisher
Elsevier
dc.subject - en
Health
Health and safety
dc.subject - fr
Santé
Santé et sécurité
dc.subject.en - en
Health
Health and safety
dc.subject.fr - fr
Santé
Santé et sécurité
dc.title - en
Gamma spectrum and coincidence summation simulations with Geant4 in the analysis of radionuclide using BEGe detector
dc.type - en
Article
dc.type - fr
Article
Download(s)

Original bundle

Now showing 1 - 1 of 1

Thumbnail image

Name: Gamma spectrum and coincidence summation simulations with Geant4 in the analysis of radionuclide using BEGe detector.pdf

Size: 1.73 MB

Format: PDF

Download file

Page details

Date modified: